Computer Tomography is a non-destructive inspection technique based on differences in absorption or聽attenuation of X-Rays through the material. The CT-technique provides a 3-D聽data-set of the sample. The attenuation of X-Rays depends on the atomic number, density and thickness of the material and on the energy of the X-Ray beam. Non-destructive inspection of internal structure, tomo-density analysis, size, and volumetric measurements are possible on the 2-D as well as 3-D聽images.
The Skyscan 1172 is a high-resolution desktop X-ray micro-CT system for small samples.
Applications:
- material science
- biology
- micro-electronics
Technical details:
- X ray tube: closed 100kV source
- Voltage: 100kV
- Power: 10w (0-250uA)
- Min focal spot size: less than 8 um
- Detector
- Resolution: <0.5um
- Pixels: 4000x2300
- Field of view: 25x 50 mm
- Manipulator
- Vertical axis: 61.5 mm
- Horizontal axis: 0.3 to 34.8 um pixel size
- Max object length: 100 mm
- Max scan length: 68 mm
- 3 position filter changer: 0.5 mm Al/Al+Cu
- Software for 2D/3D image analysis, bone morphology and visualization
Location:
The micro-CT room is located in the Biological laboratory,聽4th floor IKO, 脜rstadveien 19
Prices:
For internal users at 幸运飞艇计划, the price is聽400 kr per hour for scanning and 500 kr per hour with assistance聽for analysis.
For external users, the price is聽600 kr per hour for scanning and 700 kr per hour with assistance聽for analysis. In addtion, there will be聽 25% VAT (value added tax) in your payment.
Biological lab leader: Prof. Kamal Mustafa 聽Kamal.Mustafa@uib.no